773232AH
13-22
Troubleshooting
Cytometer Messages Table
Set Manual Mode
Status confirming samples will be run
in the MCL Manual Mode.
Select Run to continue in this mode.
To discontinue operation in this mode
depress the MCL Manual Mode button on
the Cytometer Control toolbar to return to
the MCL Automatic Mode.
Set Single Tube Mode
Status confirming samples will be run
in the Single Tube Fixed Mode.
Select Run to continue in this mode.
To discontinue operation in this mode,
depress the Single Tube Fixed Mode button
on the Cytometer Control toolbar to return
to the MCL Automatic Mode.
Sheath Cube Level Error
The 10 L sheath cube is empty and
the instrument has entered the idle
state.
Replace the external sheath fluid container.
See
Replace The 10 L External Sheath Fluid
.
Sheath Cube Level Warning
The 10 L sheath cube is empty.
Replace the external sheath fluid container.
See
Replace The 10 L External Sheath Fluid
.
Sheath Pressure Error
During acquisition, the sheath
pressure went out of range and
acquisition was stopped.
1.
Check sheath fluid container cap for
tightness.
2.
If problem continues,
3.
If problems continues, contact your local
Beckman Coulter Representative.
Sheath Pressure Warning
The sheath fluid pressure is outside
the system’s operating range.
If warning persists, see Sheath Pressure
Error.
Start of Data Decode Error
The header at the beginning of the
data that is passed from the
instrument to the workstation could
not be decoded.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
System Pressure Error
During acquisition the system
pressure was lost and the acquisition
was stopped.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative
System Pressure Warning
The system pressure is outside the
operating range.
If warning persists, see System Pressure
Error.
System Pressurization Error
Pressure line is not connected
between the power module and the
Cytometer.
Connect the pressure line.
The pressurized air supply is outside
the system’s operating range.
1.
Go to the Ready State, then check that
the system pressure is 30 psi.
2.
Run a sample and monitor the system
pressure. If the system pressure drops
below the range specified in the
instrument manual, then contact your
local Beckman Coulter Representative.
Short circuit.
Check fuses.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......