773232AH
10-105
Using Navios Software
Plots Menu
10
How is Prism Created?
In single parameter analysis of surface marker data, one or two regions are normally set to measure
both Negative and Positive cells. We can then look at a second parameter and calculate the 2
phenotypes from parameter-1 positive populations and the two phenotypes from parameter-1
negative populations (four phenotypes in total).
We can repeat this with 3, 4 or more parameters.
The Prism Divider can be used instead to mark the dividing line between the Negative and Positive
cells.
Prism Dividers
A Prism is created by setting one or more prism dividers. There are three ways of setting prism
dividers, two of them involve graphical positioning (single and dual prism dividers). The third uses
Analysis > View/Modify Prism
that can set the divider numerically on any divider. Any changes to the
divider on one plot is replicated on any other plot using that parameter.
Each new parameter that has a prism divider set on it multiplies the total number of phenotypes by
2.
Only one divider can be set per parameter.
Setting a Single-Parameter Prism Divider
Single parameter prism dividers are set by placing a
Single-parameter histogram in focus and selecting
. The divider can then be set by clicking on the
selected plot, positioning the divider, and then clicking
once more to fix the divider position.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......