773232AH
13-24
Troubleshooting
Cytometer Messages Table
Violet Laser Power Error
During acquisition, the violet laser
output power went out of range and
acquisition was stopped.
If error persists, contact your local Beckman
Coulter Representative.
Violet Laser Power Warning
The violet laser power is unstable.
If laser does not stabilize within a
reasonable time or frequently goes unstable
then restart the entire system. If error
persists, contact your local Beckman
Coulter Representative.
Waste Backpressure Error
The filter on the waste container vent
line is probably wet, clogged, or
disconnected. There is not enough
empty volume in the 20 L waste
container for further sample
analysis.
1.
Empty the 20 L waste container. See
2.
Check the waste vent filter for the
presence of liquid and a proper
connection.
3.
If problem continues, contact your local
Beckman Coulter Representative.
Waste Chamber Level Error
The liquid level in the internal waste
chamber is too high and acquisition
has been stopped.
1.
Empty the 20 L waste container. See
2.
Check the waste vent filter for the
presence of liquid and a proper
connection.
3.
Restart the entire system. If error
persists, contact your local Beckman
Coulter Representative.
Waste Chamber Level Warning
The liquid level in the internal waste
chamber is too high.
1.
Empty the 20 L waste container. See
2.
Check the waste vent filter for the
presence of liquid and a proper
connection.
3.
Restart the entire system. If error
persists, contact your local Beckman
Coulter Representative.
Waste Cube Level Error
The is not enough empty volume in
the waste container for further
sample analysis and the idle state
has been entered.
Empty the waste container. See
.
Waste Cube Level Warning
The waste container is full.
Empty the waste container. See
.
If the waste container is not full, the
waste sensor failed.
contact your local Beckman Coulter
Representative.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......