Using Navios Software
Cytometer Menu
10
Cytometer Control Parameter Setup
This screen is used to select the Signals required in the protocol. The signals that are available are
selectable by checking the options you require.
NOTE
If you add a parameter to an existing protocol, the Detector Names revert to the default names shown
below. You cannot make any changes on this screen when running a
.
CAUTION
Risk of erroneous results if you assign the same parameter name to more than one
detector. Use a different parameter name for each detector.
You can arrange the order within the
Selected Signals
list box by highlighting the appropriate
item(s) and Drag and Drop into your chosen order within the
Selected Signals
list box.
CAUTION
Risk of erroneous results if you modify a protocol by removing or changing the
order of selected parameters. Verify that the protocol's plots, regions and gate
assignments are as desired before reporting results.
To remove a previously selected item from the
Selected Signals
box, uncheck the parameter no
longer required.
Ratio
– Choose from the drop down box the appropriate option for the Numerator and Denominator
if required.
A detector name may be entered, this is displayed in all
Cytometer Control
screens but does not
appear on plots.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......