773232AH
2-3
Installation
Special Requirements
2
Power Consumption
<1500 Watts
Ambient Temperature and Humidity
Temperature fluctuations within the ambient temperature range can affect performance. For
optimal performance, Beckman Coulter recommends the ambient room temperature is maintained
between 15.5°C and 32°C (60°F to 90°F) and does not fluctuate more than
2.8°C (
5°F) from the
ambient temperature at alignment. Keep the humidity between 30% and 85%, without
condensation.
Heat Dissipation
Heat dissipation is 720W (2457 Btu/hour). Provide sufficient air conditioning (refer to Ambient
Temperature and Humidity).
Acoustic Noise Level
60db
Drainage
The waste line from the Cytometer is connected to a 20-L waste container. Dispose of the waste in
accordance with your local regulations and acceptable laboratory procedures.
Europe and
other
applicable
countries
One dedicated line at 220 Vac, 50/60 Hz
at 10 A for the Supply Cart.
or
One dedicated line at 240 Vac, 50/60 Hz
at 10 A for the Supply Cart
Two non-dedicated lines at 220 Vac,
50/60 Hz at 10 A - one for the tower
computer and a second, for the
monitor. A third non-dedicated line
is required if the optional printer is
being installed.
Japan
One dedicated line at 100 Vac, 50/60 Hz
at 15 A for the Supply Cart.
Three non-dedicated lines at
100 Vac, 50/60 Hz at 15 A - one for
the tower computer and a second,
for the monitor. A third non-
dedicated line is required if the
optional printer is being installed.
Country
Dedicated Lines with Isolated Grounds
Non-Dedicated Lines
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......