773232AH
10-160
Using Navios Software
Cytometer Menu
the Dots field to the number of Dots required. If a zero is entered, an error message displays showing
range of 10 - 9999. The Dots value is stored in the Protocol
QuickCOMP Mode
QuickSET Mode
Baseline Offset
Enable this checkbox to switch on Baseline Offset. Enabling this checkbox will override the Baseline
Offset selections on the Workspace Preferences - Plot Display tab. Further information concerning
the setting of this option can be found under
Workspace Preferences - Plot Display
.
E
nabling
QuickCOMP
dialog box displays red
slider bars on all dual fluorescence plots.
These sliders are used to intuitively adjust the
compensation coefficients and update the
compensation values in the
Clicking the arrows adjusts the settings by
±0.1%, clicking either side of the slider adjusts by
±1.0% or drag the slider to place the cells in the
required position.
When
QuickCOMP
is enabled
QuickSET
mode is
not available.
NOTE
Listmode Compensation must be performed
using the runtime protocol or a protocol with
equivalent parameters as the runtime protocol.
Enabling
QuickSET
on the
dialog box displays
slider bars on all plots.
These sliders are used to adjust voltages at
the data plots and update the high voltages
in the
Cytometer Control Acquisition Setup
.
Clicking the arrows adjusts the voltage by
±0.1 volt, clicking either side of the slider
adjusts by ±1.0 volts or drag the slider to
place the cells in the required position.
When
QuickSET
is enabled
QuickCOMP
mode
is not available.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......