Using Navios Software
Cytometer Menu
When in Pause mode, the Acquisition time on the status bar does not continue to increment
although the Elapsed time does continue to increment. The total Acquisition/Elapsed time can also
be viewed from the FCS Keywords associated with the beginning time and end time of acquisition.
Number of Events
The current total number of events acquired is displayed here.
When acquiring, this value increments to the maximum events set or until one of the Stop
conditions has been met. See
When acquiring data in Setup Mode, this item cycles in the range from 0 to 10 times the number of
Events shown in the Dots Displayed. The value appears to wraparound, that is, if you have requested
5,000 dots to be displayed, this item increments to 4,999. On the 5,000th event the total events
counter resets to 0.
This ensures that you can continue to display data without reaching a set stop condition.
Cytometer Error/Warning Messages
Error/Warning messages generated by the Cytometer. The software translates any received error
codes and displays the appropriate text in the status bar. When there are no messages coming from
the instrument, this field defaults to "No messages Received". Double clicking this field brings up
the Cytometer Status Messages dialog. See
in the Troubleshooting chapter.
Flow Rate
Windows Status Bar
The Windows Status Bar normally gives a brief description of selected menu items.
When drawing or editing Regions, the Status Bar shows Channel for X and Y parameters
respectively as they are drawn.
This item shows the relative flow rate of the
Cytometer during acquisition. Click on the drop
down box and choose the required rate.
The Flow Rate can only be displayed as Low,
Medium or High. The Flow Rates are
approximately: 10, 30 and 60
μ
L/min respectively.
NOTE
_STAND protocols are hard-coded to a Flow
Rate of Medium. This flow rate is not passed to
remaining tubes in the AutoSetup Application.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......