773232AH
3-12
Operation Principles
Parameters
Figure 3.10
Time vs Fluorescence plot
The minimum stop time is 10 seconds, the maximum stop time (maximum duration) is
1,200 seconds and the default stop time is 300 seconds (5 minutes).
When you assign the TIME parameter to a plot axis, the divisions on the axis change accordingly.
To find the time (in seconds) per channel in a one-parameter histogram, divide the stop time (in
seconds) by 1,024 (0.001 second = 1.0 ms).
For a two-parameter plot, divide by 64, 128, 256 or 512 depending upon the plot resolution you are
using.
RATIO Parameter
The RATIO parameter is calculated, not acquired directly. When you select a parameter, you specify
which signal is the numerator and which is the denominator.
A ratio of 1 is at channel 1,023. If you assign RATIO to a plot axis, RATIO events appear at a lower
channel if the intensity of the numerator signal is less than the denominator signal.
To calculate the actual ratio at a particular intensity for a one-parameter histogram, divide the
intensity by 1,024. For a two-parameter plot, divide by 64, 128, 256 or 512 depending upon the plot
resolution you are using.
RATIO
Numerator
Denominator
------------------------------------
1024
×
=
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......