Troubleshooting
Software Messages
13
Software Messages
Navios Software Messages
lists the Navios software messages, with their cause and what to do about them. These are
messages produced by the Software.
contact your local Beckman Coulter Representative if:
•
The recommended action does not solve the problem.
•
You need help.
Table 13.3
Software Messages
Message
Probable Cause
Recommneded Action
Unspecified error
An unknown error occurred.
Contact your local Beckman Coulter
representative.
Nothing to modify!
Modify button pressed with no dies
listed.
Select Dye
Unable to read Error Code Strings
Failed to read error code string
Reboot the computer, if problem persists
contact your local Beckman Coulter
representative.
The file contains errors. The data
checksum does not match the data.
Listmode data checksum stored does
not match checksum re-calculated.
The listmode data is corrupt and the you
should recover the original listmode file
from backup or archive.
Baseline Offset - not available
On starting software the 3mv binary
file is not found.
Reboot the computer, if problem persists
contact your local Beckman Coulter
representative.
Multiplexor Address of new labels
does not match existing parameters
Continue?
Parameter label does not match,
asks you to continue.
Reboot the computer, if problem persists
contact your local Beckman Coulter
representative.
Missing NoOfTest field in file
<<file-name>>
On inserting a panel into the
listmode playback it is discovered
that the number of tests information
is missing from the file.
Recreate the panel file.
Missing NoOfPanels field in file
<<file-name>>
Number of panels information
missing from the worklist.
Recreate the worklist.
Missing NoOfTest field in section
Panel <<panel-number>> Info in
file <<file-name>>
Number of tests information missing
from the panel specified in the
worklist.
Recreate the panel file.
The BCAP service does not appear to
be running.\nSoftware cannot start
without BCAP.
Starting software without BCAP
causes this error message.
Reboot the computer, if problem persists
contact your local Beckman Coulter
representative.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......