Cleaning Procedures
Clean The Sampling System
11
When to Clean the Sampling System
Routine and Sample Head Cleaning Procedures
Perform both the routine and the sample head/probe cleaning procedures before you perform Daily
and:
•
When you change laboratory application procedures, especially if you are using vital
fluorescent stains. If vital stains such as propidium iodide, ethidium bromide, acridine orange,
thiazole orange, Coriphosphine-O, Fura 3, or fluorescein diacetate, are used, perform these
cleaning procedures immediately after using the dyes.
•
Immediately prior to running any immunophenotyping application if vital stains are being used
on the same instrument.
•
When you observe a significant increase in debris or background counts.
Clean The MCL Sample Head And The Sample Probe
To remove any crystal or debris buildup, perform this procedure as often as needed. See also,
.
1
.
2
Lift up the MCL cover.
NOTE
If a carousel is present, remove it.
CARO
USEL
CLE
ANIN
G:
PREP
ARE
A SO
LUTI
ON O
F 1 P
ART
HOU
SHO
LD B
LEAC
H (5%
SOLU
TION
OF S
ODIU
M HY
POC
HLOR
ITE)
AND
9 PAR
TS
WATE
R, RI
NSE
CARO
USEL
WITH
SOL
UTIO
N,
LIGH
TLY R
UB IF
NEC
ESSA
RY.
IMM
EDIA
TELY
RINS
E WIT
H
CLE
AR W
ATER
CARO
USEL
CLE
ANIN
G:
PREP
ARE
A SO
LUTI
ON O
F 1 P
ART
HOU
SHO
LD B
LEAC
H (5%
SOLU
TION
OF S
ODIU
M HY
POC
HLOR
ITE)
AND
9 PAR
TS
WATE
R, RI
NSE
CARO
USEL
WITH
SOL
UTIO
N,
LIGH
TLY R
UB IF
NEC
ESSA
RY.
IMM
EDIA
TELY
RINS
E WIT
H
CLE
AR W
ATER
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......