Quality Control
Autosetup II Wizard
7
e.
If the HPCVs pass,
on the Autosetup Wizard or
to continue with Additional scheduled QC applications.
f.
If the HPCVs are not within the upper limits identified in the protocol's region name, you
may need to
after performing one or more of the following
troubleshooting steps:
1)
Ensure the Flow Rate is LOW
2)
the system
g.
If the HPCVs are still not within the upper limits, select Abort and perform the following
troubleshooting steps and repeat scheduling and running the alignment application.
1)
Run a
Cycle
2)
3
STANDARIZATION APPLICATIONS
When the _STAND protocol's data acquisition starts, view all histograms to ensure peaks fall
within regions.
NOTE
When running a negative control verify that the AutoGate is around the population of interest in
the FS vs SS dotplot. If you need to, adjust populations.
a.
Auto
Adjust
Disabled
as soon as the AutoSetup wizard displays.
b.
Adjust voltages to place the lymphocytes in a suitable position.
c.
Adjust voltages to place the fluorescence populations within the first log decade.
d.
.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......