773232AH
5-31
System Overview
Verifying A Worklist
5
3
After all the information has been added, you can use
to change the order.
Worklist Panel
1
.
2
Select which panel you wish to open and
.
3
After all the information has been added, you can use
to change the order.
You can use this option to add another Panel to an existing Worklist.
Worklist Test
to insert a single Row at the end of the current Panel.
NOTE
A duplicate of the first protocol in the panel is added to the current panel. If a different protocol is
required,
in the protocol field.
Verifying A Worklist
Always verify carousel and tube location fields before starting acquisition and ensure that the
Acquisition Manager information accurately reflects the tubes in the carousel being run. The order
of the tubes in the carousel must match the order of the tubes in the Worklist.
The default protocol (Ctrl N) must be saved prior to being used in the Worklist. If the default
protocol is not saved with a specific name prior to using it, that tube may be run with unexpected
protocol settings depending on the Save Protocol selection in Workspace Preferences. Also, the
default protocol name will default to "setup.def" in any associated panel template and have no
regions.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......