773232AH
13-15
Troubleshooting
Cytometer Messages Table
13
Change Font
To change the font:
Format > Font
, and select the new font.
Find Help Topics
To find Help on other topics:
Help > Help Topics
.
Cytometerarchive.log File
T
he cytometerarchive.log file contains messages that were automatically moved from the
cytometer.log file after 30 days. This log is for Service use.
Cytometer Messages Table
lists the messages in alphabetical order, with their cause and what to do about them.
These are the messages produced by the Cytometer.
contact your local Beckman Coulter Representative if:
•
The recommended action does not solve the problem.
•
You need help.
Table 13.2
Cytometer Messages
Message
Probable Cause
Recommended Action
Acquisition Yield Warning
The threshold events exceeds the
processed and displayed events.
1.
Dilute the sample .
2.
Change the discriminator setting.
3.
Change the sample flow rate to medium
or low.
4.
Restart the entire system. If error
persists, contact your local Beckman
Coulter Representative.
Ambient Sensor Error
The ambient temperature sensor
failed during acquisition and the
acquisition was stopped.
If error persists, contact your local Beckman
Coulter Representative.
Ambient Sensor Warning
There has been no change in the
measured ambient temperature for
some time and the sensor may have
failed.
If warning persists, contact your local
Beckman Coulter Representative.
Ambient Temperature Warning
Ambient Temperature is outside the
system's specifications.
Reduce or wait till ambient temperature is
within specification.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......