773232AH
12-39
Replace/Adjust Procedures
Replace An Optical Filter
12
Remove Filter Holder
1
Wear powder-free gloves to perform this procedure.
2
Open the Front Cover.
3
Loosen the four thumbscrews on the filter array cover and remove it.
NOTE
You need to unscrew the upper left thumbscrew with a screwdriver.
CAROU
SEL CL
EANING
:
PREPA
RE A S
OLUTIO
N OF 1
PART H
OUSHO
LD BLE
ACH (5
%
SOLUT
IONOF
SODIUM
HYPOC
HLORIT
E) AND
9 PART
S
WATER
, RINSE
CAROU
SEL WI
TH SOL
UTION,
LIGHTL
Y RUB
IF NEC
ESSAR
Y.
IMMED
IATELY
RINSE
WITH
CLEAR
WATER
CAROU
SEL CL
EANING
:
PREPA
RE A S
OLUTIO
N OF 1
PART H
OUSHO
LD BLE
ACH (5
%
SOLUT
IONOF
SODIUM
HYPOC
HLORIT
E) AND
9 PART
S
WATER
, RINSE
CAROU
SEL WI
TH SOL
UTION,
LIGHTL
Y RUB
IF NEC
ESSAR
Y.
IMMED
IATELY
RINSE
WITH
CLEAR
WATER
1
3
5
6
8
9
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......