7-2
Quality Control
QC Processes
Discrimination Parameter
The various parameters produce differences in signals ranging from robust to weak signals which
in turn affect the amount of noise in the baseline. The discriminator is applied to the parameter of
your choice, based on your application.
Discriminator Value
A threshold value is applied to the selected parameter signal to control the amount of background
that is measured. Setting the discriminator too low could include a higher amount of debris while
a setting that is too high might eliminate the desired events. See
, for information on
selecting and modifying discriminators.
Flow Rate
The flow rate governs the events per second that pass by the flow cell. The concentration of the
particle of interest should be considered when selecting a flow rate so that coincidence (doublets)
is minimized. Generally, the faster the flow rate, the higher the probability for creating coincidence
and higher CVs (broadening of the peak). See
, for instructions on changing the flow rate.
section for an example of using the Time of Flight
parameter to detect coincidence.
Voltages and Gains
These settings are used to position the signal for each parameter in the desired position. See
Cytometer Control Acquisition Setup Tab
for instructions on changing signal voltages and gains.
Plot Selection
There are several plot types available to view data: bi-variate (dot plots/density plots) and
histograms. Ability to view data to affirm appropriate cytosettings (including voltage, gain and
discriminator assignment) is determined by parameter assignments to the selected plots. See
, for instructions on creating plots.
Regions and Gating
Regions and gates allow for statistical information to be generated from the data set. Creation and
placement or regions and gate assignments must be verified for the dataset. See
, for
instructions on creating and modifying regions and gates.
QC Processes
The chart below shows which QC materials are needed for each QC process.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......