773232AH
10-154
Using Navios Software
Flowpage Menu
2
the
button. All items in the group shrink to the size of the item which has the smallest
area.
Align Left
1
Select the items to be aligned. See
Select a Group of Items Within FlowPAGE
.
2
. All items are now aligned to the left of the last selected (grey handles) item.
Align Right
1
Select the items you wish to align. See
Select a Group of Items Within FlowPAGE
2
. All items are now aligned to the right of the last selected (grey handles) item.
Align Top
1
Select the items to be changed. See
Select a Group of Items Within FlowPAGE
2
. All items are now aligned to the top of the last selected (grey handles) item.
Align Bottom
1
Select the items to be changed. See
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......