Using Navios Software
Analysis Menu
10
Log-Lin
Selecting the Log-Lin mode calculates the mean values by determining the arithmetic mean of the
raw ADC channels and then converting this to its relative linear channel equivalent.
Automatic Statistic Percentiles
This is the channel, which divides the events in a region by the percentile value. The default
percentile is 50 (the median) but can be set to request any desired value. Different values may be set
for both the X and Y axis.
Absolute Count Calibration
Use absolute count calibration particles for the calculation of absolute numbers of cells present in
a sample.
As a sample is analyzed, the numbers of a particular cell type in a region, and the numbers of
absolute count calibration particles present in a CAL region are compared, and the absolute
concentration of the cell type is then calculated automatically.
A listmode files replays with its runtime Cal Factor. Use this screen to enter a different Cal Factor to
calculate individual listmode absolute counts.
NOTE
•
A CAL region MUST be set for the system to be able to calculate an absolute count.
See
•
Listmode replay uses the runtime protocol Cal Factor unless you select a different Cal Factor before
replaying the file.
To display the absolute count,
Cells/
μ
L
from the Analysis/
menu option.
Setting a CAL Factor for Absolute Counts
From the list of
Available Absolute Count Calibrators
,
select the Lot Number currently in use.
Enter a
Manual Absolute Count Calibration Value
if the
required Lot is not present in the list. A manual
calibration value is only used for the current session, it
cannot be saved. A listmode file replays with its
runtime Cal Factor.
NOTE
The number of events in the CAL region must be
greater than 1000, otherwise absolute counts are not
determined.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......