773232AH
3-11
Operation Principles
Automated Software Features
3
causing Published data to be formatted as a single row of values. In the case that the Publish Data to
MS Excel option is selected, this progressively builds a table on a single spreadsheet.
Cleanse Protocols: Any Protocol name beginning with the word “Cleanse” overrides the Output
Options setting for Save LMD on the
Workspace Preferences - Acquisition Options
tab. This allows
you to create protocols that do not write a listmode file to disk on completion.
AS Protocols: There is a series of special protocols which include “AS” in their names. These are
components of the AutoSetup applications which operate with the Auto Setup Wizard to set up
Volts, Gains and Compensation.
Special Panels
The cleanse panel utilizes two cleanse protocols (‘cleanse bleach.PRO’ and ‘cleanse di water.PRO)
that you should use at the end of the working day or shift on your Navios flow cytometer.
Automated Software Features
Navios software contains the following automated software features.
•
•
•
•
•
•
•
Elliptical and Contour AutoGating
•
Parameters
TIME Parameter
The TIME parameter is the amount of time, in seconds, the instrument acquires data. It is displayed
on the plot axis in 1-second resolution. The axis labels vary, depending on plot resolution and stop
time (duration).
NOTE
Including TIME as a selected parameter in all protocols allows for an internal quality check on the data
acquisition. A TIME versus fluorescence plot may be helpful to monitor system fluidic and optic conditions
during acquisition of any given sample. Monitor consistent fluorescence over time as shown below.
Unexpected fluctuations in the pattern of fluorescence may indicate compromised fluidics or optic
conditions.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......