773232AH
4-33
Specifications
Performance Characteristics And Specifications
4
Performance Characteristics And Specifications
NOTE
Unless stated otherwise, all performance characteristics and specifications were achieved at a
medium flow rate.
Carryover
Particle carryover is <0.1% from one specimen to another when the number of gated events is
10,000.
Fluorescence carryover on the Navios system was assessed by analyzing an unstained control
sample after the acquisition of a sample stained with a vital dye. Following the acquisition of three
tubes of Cyto-trol Control Cells stained with 20ul of Acridine Orange, the system was cleaned
according to the
. Subsequently, 3 tubes of unstained Cyto-trol Control
Cells were analyzed. The average mean channel fluorescence shift obtained in the unstained sample
acquisitions was less than 1%.
Acquisition Rate
The Navios was verified to analyze at least 200,000 events from FS, SS, FITC, PE, ECD, PC5.5, PC7, and
APC parameters with event rates from 1384 to 40822 events per second. Electronic pulses were
counted from the beads passing through the aperture as measured by discriminator events. The
count from the histogram display/printout was assessed against the number of pulses counted
(electronic count). Yield percentage was taken from the Stored Events counter. At 25,000 events per
second, the yield was measured to be 90%.
Data Acquisition Throughput
Throughput, including printing, of 80 tubes/hour of 10,000 normal leukocytes (~2,500
lymphocytes/uL) was obtained at a high flow rate. The throughput of a concentrated sample was 88
tubes/hour was obtained at 10,000 events per second when the stop count was set to 100,000 gated
events.
Acquisition throughput is dependant on many variables, including flow rate, data rate, number of
parameters collected, gated events, data plots, FlowPAGE reports and printouts generated. Changing
these variables, such as decreasing the number of events collected, increasing sample concentration
or increasing the flow rate can increase your acquisition throughput; while increasing the number
of events collected or running at a lower data rate can decrease your acquisition throughput.
Precision for Surface Markers
See reagent package insert for precision specifications of surface markers.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......