773232AH
4-34
Specifications
Performance Characteristics And Specifications
Scatter Resolution
Scatter resolution on the Navios Flow Cytometer was measured using 0.404
μ
m particles from
Thermo Scientific. The noise was set to the bottom of the scale, using FS HV = 700, Gain at 200,
Discriminator at 47 and setting the Forward Scatter detector to the Wide2 setting. The Navios was
able to clearly show baseline resolution between the noise and the 0.404
μ
m diameter particles.
Forward Scatter
The HPCV of the integral signal intensity values using Flow-Check Pro fluorospheres is <2% from the
blue laser.
Fluorescence
The HPCV of the integral signal intensity values using Flow-Check Pro fluorospheres is <2% for FL1-
FL4 and <2.5% for FL5 from the blue laser, <3.0% for FL6-FL8 from the red laser and <4.0% for FL9-
FL10 from the violet laser.
Sensitivity
Fluorescence
Three measurements of multi-level fluorescence sensitivity particles were taken on 3 Navios Flow
Cytometers. Acquisition was conducted on 10,000 bead stop count at the medium flow rate. For the
FITC and PE measurements, 8 peak Spherotech Rainbow Calibration Particles (RCP-30-5A) were
used. For the PeCY5, and APC, Spherotech Ultra Rainbow Calibration Particles (URCP-38-2K) were
used. The following values represent the average of the measurements taken.
•
<112 MESF for FITC
•
<78 MESF for PE
•
<15 MESF for PC5
•
<75 MESF for APC
Laser Characteristics
Day-To-Day
The mean channel value of the integral signal intensity of alignment verification fluorospheres
from the standard PMT’s off the blue, red and violet lasers does not vary more than ±5% from any
integral signal channel number obtained over a period of 8 days.
Within Day
The mean channel value of the integral signal intensity of alignment verification fluorospheres
from the standard PMT’s off the blue red, and violet lasers does not vary more than ±5% from any
integral signal channel number obtained within a period of 24 hours.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......