773232AH
8-9
Running Samples
Running Samples - MCL Manual Mode
8
6
or
to select the Worklist, Panel or Protocol
you want to use.
NOTE
If you load a Panel or Worklist that contains a deleted Protocol, the software prompts you to use
Wizard to correct the problem. Panel names for the corrected protocols are not
displayed in the Worklist however they are displayed on the Panels tab.
7
Enter the carousel ID number in the Acquisition Manager.
8
to place the Cytometer in MCL Manual Mode.
9
to begin processing samples.
10
Enter the tube location to process and
.
During the sample cycle the following series of Cytometer status messages appears:
Awaiting Sample
Preparing Sample
Acquiring
Stopping
.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......