773232AH
10-193
Using Navios Software
Creating And Importing Panel Report Templates
10
1
The equation result is selected from a drop-down list that includes:
%&#
(Both) = Percent result is calculated from percents and is placed in the Result column and
Count result is calculated from counts and is placed in the Count column.
Percent
= Result of the equation is placed in the Result column with a % sign.
Count
= Result of the equation is placed in the Count column.
Result
= Result of the equation is placed in the Result column, no % sign.
See explanation for
(5)
for additional equation options.
CAUTION
Risk of misleading results if you do not specify the correct cell population. Specify the
correct cell population even if you do not want to print the line since other lines may refer
to this line.
2
When selected, lists all the protocols in the panel. From the drop down list select the protocol to
extract results from.
3
Text description of the current template line
4
Displays the optional statistics for the result that was selected in the Region field in the Protocol area
of the Local tab. Select up to two of the following for each Results Row: X-Mean, Y-Mean, X-Mode,
Y-Mode, X-Median, Y-Median, X-CV, Y-CV, X-HPCV, Y-HPCV. Only statistics that were selected in the
protocol during acquisition are available as options. Optional statistics cannot be selected when
Equation is selected.
5
Use one of the following methods to define a custom equation.
•
Type an equation using the (), *, /, +, -, @, & operators. Result Rows are designated in the
equation as R001 to R999.
•
The Row number can be preceded by,
"
%
=
Use the value in the Result column of the row
"
#
= Use the value in the Count column of the row
"
@
= Use the value in the Opt Stat 1 column of the row
"
&
= Use the value in the Opt Stat 2 column of the row
•
Define an equation by clicking on a cell in the Results Section to add it to the equation. For
example, clicking on the Count column of Row 5 adds #R005 to the equation.
Right clicking on this field displays the following options,
+
Add
-
Subtract
*
Multiply
/
Divide
DF
= Dilution Factor
HV
= Harvest Volume
BW
= Body Weight
6
Select a cell population (LY, MO, NE, EO, BA, Other, WBC, RBC, PLT, or blank) based on the gating
that was used. The available selections correspond to the hematology information in the database.
A selection must be made when the calculation method is
Cells/
μ
L HEM
; otherwise, Cell Pop is
grayed but retains its value. Cell Pop is blank when Equation is selected.
7
Adds a row using the information entered and blanks all of the fields.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......