773232AH
5-18
System Overview
Creating Regions
1
Highlight a dual parameter plot.
2
.
3
A Rectangular Region cursor
appears in the highlighted plot.
4
Position the cursor to the point where you want the Region to begin.
5
Click and release the mouse button. A Rectangular Region start box is displayed on screen. By
moving the mouse you can increase and decrease the size of the Rectangular Region. When the
desired Rectangular Region appears, click the mouse button to complete the Region.
6
The Region Name appears on the plot.
Create Quadrant Regions
The Quadrant Region is a FlexQuad Quadrant Region. You can adjust the angle of quadrant dividers
to adequately analyze populations that are properly compensated with the negative data display.
The dividers between FlexQuad quadrants 1 and 2 and between quadrants 2 and 4 are 'flexible'
which allows you to better analyze broadly spread populations.
1
Highlight a dual parameter plot.
2
.
3
A crosshair cursor
is displayed in the current window.
4
Position the cursor on the plot where you want the intersection point of the region and
to show the quadrant lines (cross hair).
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......