773232AH
10-51
Using Navios Software
File Menu
10
Dot Plot Coloring Mode
This alternates Color Dot plot mode between Color Blend and Color Precedence modes:
•
Color Blend Mode
•
Color Precedence Mode (default)
AutoGating Sensitivity
Select the sensitivity of the AutoGate region drawn.
Elliptical Sensitivity
Settings are 0.11, 0.33 and 1.1 (% of processed peak). Selecting 0.11 creates a larger region, whereas
selecting 1.1 creates a smaller radius region around the target population.
Contour Sensitivity
Settings are 1, 2, 3, 4, and 5. Level 1 = 0.11% processed peak at a resolution of 64x64, levels 2, 3 and
4 are 0.11, 0.33 and 1.1% at a resolution of 128x128. Level 5 is 1.1% processed peak at a resolution of
256x256.
Contour Travel
Input a maximum value the contour AutoGate region travels from the mean. The Minimum value is
25; Maximum is 250 channels. Use this option to slightly adjust the AutoGate from tube to tube if
slight changes in lysis cause the target population to move slightly. This option is not available for
elliptical AutoGates.
AutoGating Failure
You can select an error option when the AutoGate algorithm fails during acquisition in the MCL
automatic mode.
Pause on Error
The system pauses the current acquisition when an AutoGating error occurs and awaits further
input before proceeding.
Abort On Error
The system automatically aborts the Worklist when an AutoGating error occurs.
Ignore Errors
The system continues processing the Worklist when an AutoGating error occurs.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......