773232AH
13-17
Troubleshooting
Cytometer Messages Table
13
Carousel Label Read Warning
The first attempt to read the carousel
bar-code label failed.
If warning persists, see Carousel Label read
Error.
Carousel Rotate Warning
The carousel failed to rotate to the
correct position after several
attempts.
1.
Check that a carousel is in place in the
MCL.
2.
Check that there is no obvious
obstruction (sample tube) in the MCL
area.
3.
Check that the 30 psi gauge on the
Pneumatic Supply is okay.
4.
Try using another carousel.
5.
If no obstruction is found and pressure
okay, contact your local Beckman
Coulter Representative.
Cleanse Level Error
The requested cleanse procedure
could not be performed due to the
low level of cleaning agent.
Fill cleaning agent container. Refer to
.
Cleanse sensor failed if the cleaning
agent container is full but this error
message displayed.
contact your local Beckman Coulter
Representative.
Cleanse Level Warning
Low cleaning agent.
Fill cleaning agent container. Refer to
.
Cleanse sensor failed if the cleaning
agent container is full but this error
message displayed.
contact your local Beckman Coulter
Representative.
Control Socket Error
A control communication error has
occurred between the workstation
and the instrument.
Check the Ethernet cable between the
workstation and the cytometer. Restart the
entire system. If error persists, contact your
local Beckman Coulter Representative.
Corrupt Configuration Error
The instruments configuration
information is corrupted.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Data Communication Lost Error
Data socket connectivity has been
lost.
Close the workstation software and open it
again. If problem persists, contact your local
Beckman Coulter Representative.
Data Rate Warning
The discriminated data rate is higher
than can be handled by the
instrument and data is being lost.
1.
Dilute the sample or change the
discriminator setting.
2.
Check that the sheath fluid container
cap is tightened.
3.
Change the sample flow rate to medium
or low.
4.
If problem continues,
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......