773232AH
8-15
Running Samples
Data Review
8
Data Review
Review Data Displays
CAUTION
There is a risk of reporting incorrect results. Data displays used to arrive at the
test result for light scatter patterns, antibody staining profiles, and all gates and
boundaries should be reviewed by a lab
oratory professional when interpreting
the data. If results are suspect, follow your laboratory's procedures to resolve.
Data review should include the following:
•
Review the light scatter patterns.
—
Verify any population of interest is above the discriminator to ensure no cell loss.
—
Verify any population of interest is separated as much as possible from other populations
that are present (within biological constraints).
•
Review the antibody staining patterns.
—
Verify overall expected staining patterns are observed; in other words, verify the
fluorescence staining correlates to the antibody used in sample preparation.
—
Verify abnormally dim or negative sample staining patterns are in agreement with clinical
presentation and other diagnostic data.
•
Review Gate and Region boundaries.
—
Verify gates and regions encompass any populations of interest while at the same time
exclude undesired events.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......