773232AH
3-8
Operation Principles
Signal Processing
Integral Signal
Because the total fluorescence in all three cells is the same, but the distribution is different, the
pulse can be integrated to produce an integral signal (see
).
The height of the integral pulse is proportional to the total fluorescence and is obtained when the
cell exits the laser beam. The pulse height, however, represents the most intense amount of
fluorescence produced. The area under the pulse is proportional to the total fluorescence.
Figure 3.8
Integral and Peak Pulses
1.
Direction of sample flow
2.
Cell in laser beam
3.
Peak pulses
4.
Volts
5.
Time
6.
Integral pulses
7.
Volts
8.
Time
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......