773232AH
A-6
Bar-code Specifications
Mcl Bar-code Reader
Sample misidentification can occur from the use of incorrect, poor quality, damaged, dirty or
improperly placed bar-code labels. Follow the specifications in this section to create your bar-code
labels to prevent incorrect sample identification. See also
Putting a Bar-Code Label on a Sample
.
The instrument supports preprinted labels.
Bar-Code Label Optical Characteristics at 670 nm ±10%
•
Print Contrast Signal (PCS): 80% minimum.
•
Reflectivity of Media (RW): 80% minimum.
•
Reflectivity of Ink (RB): 16% maximum.
•
No spots or voids; no ink smearing.
•
Edge roughness is included in the bar and space tolerances.
Mcl Bar-code Reader
The MCL uses a visible-laser type reader containing a Class II laser, operating at 670 nm, with a
maximum power output of 1 mW.
Bar-code Decoder
The MCL sends a “GS” ASCII character (hexadecimal 1D) to the decoder to start operation.
Table A.2
Code-Related Specifications
Code
Interleaved
2-of-5*
Codabar*
Code 39*
Code 128*
Narrow element (NE) width
0.010 in. ±0.001
0.010 in. ±0.001
0.010 in. ±0.001 0.010 in. ±0.001
Wide element/narrow
element ratio (WE/NE)
3:1
N/A
3:1
N/A
Intercharacter gap
No
0.010 in. minimum
_NE
No
Data digits
14**
1 to 10**
1 to 7**
2 to 16
* See AIM Uniform
Symbology Specification,
Rev. 1995 for detailed
specification.
** Includes checksum
character
PCS
RW RB
+
RW
-------------------------
100%
×
=
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......