773232AH
6-11
Daily Routine
Extended Shutdown
6
7
Pour 1 L of distilled water into the internal sheath fluid container. Do NOT fill this container. A
partially filled container triggers water to be pumped from the external container of distilled
water into the lines during this procedure.
8
Fill the cleaning agent container with distilled water.
9
Cap and replace the internal sheath fluid and cleaning agent containers.
10
Perform the
except use tap water in all four tubes. Do not use any
bleach or IsoFlow sheath fluid.
11
When you start up the instrument for the first time after the extended shutdown,
•
If disconnected, reconnect the pickup tube assembly and tubing to the instrument.
•
Replace The 10 L External Sheath Fluid Container
1 L
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......