773232AH
10-130
Using Navios Software
Analysis Menu
Linear Region Properties
Lower Limit and Upper Limit
Set the region to defined limits by entering upper and lower limits for the region. These limits can
be viewed in the statistics box for each plot by selecting
Analysis >
Min/Max
.
Target Region for Automatic Setup
Required only when a multiple standardization bead preparation is used in the _STAND protocol
containing multiple regions in the Forward Scatter plot. This feature is not used when a single bead
product is used to standardize the system.
Positives Analysis
This option moves the region to the selected percent result. Specify a positive percentage value
from 0.1 to 99.9. If the positive analysis option is not selected, the percentile is unavailable. Use this
option to automatically set the region position for a negative control sample.
Interactive Linear Region Editing
When a Region's position is changed, any Gates that use the Region are automatically updated and
each plot is redrawn to reflect the changes in the Gate. When the Region is deleted any Gate which
used the Region is also automatically deleted. When a Gate has been deleted, any plot that used the
deleted Gate is reset to Ungated. Navios software allows you to resize an already created Region.
Status Bar (Linear)
The status bar normally gives a brief description of selected menu items.
Target Region for Automatic Setup
Automatic Region Name Positioning
Region statistics exported for QC
NOTE
You cannot make any changes on this screen when
running a
locked protocol other than selecting a
QC
product or choosing to export QC statistics.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......