Using Navios Software
Cytometer Menu
Cancel
Pressing Cancel will close the dialog without updating the Acquisition or Elapsed Time settings in
the Cytometer Control dialog.
Maximum Events
This is the maximum number of events, which may be acquired on the next Acquisition. To change
this, enter the new number required in the input field. However, if the Duration is reached before
the maximum number of events acquisition ceases at this point.
Live Gate
This displays all Gates that are available for the Live Gating of data during acquisition. The primary
use of a Live Gate is for the intentional removal of unwanted events, such as debris. The data outside
of the Live Gate is NOT STORED to the final listmode file by gating out all unnecessary events, thus
reducing the data set size.
The software was also designed to enable you to adjust the Live Gate for cell population drift during
long acquisition times; this does not automatically refresh the events. The listmode file will include
events that are a combination of the original and modified gate(s) used throughout the Live Gate
acquisition.
CAUTION
Improper use of the Live Gate region may generate erroneous but credible results
if the Live Gate region is modified during acquisition. If a Live Gate region is
modified during acquisition the data set will be a combination of the original and
modified gate(s). If this is not desired, then a restart must be invoked manually.
•
Ensure that the Live Gate is correctly positioned on the population(s) of interest prior to the
acquisition of data. Do not make changes to the Live Gate region or assignment during
acquisition.
•
Make changes to Live Gate regions only after activating the SETUP option in the Cytometer
Control screen resulting in a refresh of the data. If not in setup mode when changes are made
to a Live Gate region, a manual restart is required.
•
Review all data plots and results prior to reporting results.
Max Acquirable Events
A system preset limit of the Maximum Acquirable Events based on the number of parameters
selected in the protocol.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......