773232AH
7-12
Quality Control
Autosetup II Wizard
4
to print verification tube results and advance to the approve screen.
CAUTION
Cytosettings may not be saved if you clear the worklist prior to selecting Finish
and Approve in the AutoSetup II Wizard. Do not clear the worklist before Finish is
selected on the AutoSetup II Wizard and the desired Approve Instrument Settings
option in Step 5 has been selected.
5
Check the printed verification tube results to ensure instrument settings are correct and
.
NOTE
Due to differences between individual instruments, settings files should not be transferred from
one instrument to another. If you attempt to use settings files from another instrument, the software
displays
Incorrect Cytometer Serial Number.
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......