773232AH
10-159
Using Navios Software
Cytometer Menu
10
Discriminators
Lists the currently selected parameter and allows the setting and modification of discriminator
values. To change a value, select the discriminator you wish to alter and use the Slide Control on the
right hand part of the display to change the value as appropriate.
Volts
Select the appropriate Voltage for the particular Detector. This is done by using the Slide Bar on the
right side of the screen and adjusting it to the required value. For fine adjustment, the Up and Down
arrows at the ends of the Slide Bar can be used.
Gain
Adjusts the Gain amplifier. Set the Gain to the appropriate setting you require for that Detector. This
is chosen from the drop down box, which appears when you click in the box.
Details
dialog.
NOTE
Do not use an AutoGate as a Live Gate. A Live Gate should be exclusionary.
Setup Mode
Enable this checkbox to allow the continuous acquisition of data from the Cytometer. In this mode,
Stops are disabled, Printing is disabled and listmode files are not stored.
Setup Mode allows continuous acquisition of data from the Cytometer giving continuous real-time
feedback of instrument performance on the displayed plots. Any Stop & Save values set are ignored
in this mode.
Display of data continues until
Abort
is selected, to end data acquisition, or the
Setup Mode
box is
unchecked, to begin normal data acquisition.
In Setup Mode the plots are updated real-time with only the most recently acquired events
displayed. The incoming data is not saved. During Setup Mode, older events are cleared from the
screen and from memory allowing the Cytometer settings to be adjusted in real-time without
reaching a set stop condition.
Data is displayed on a “first in; first out” basis. The number of events to be displayed can be changed
by entering the required value into the
input box.
Once you have finished using Setup Mode click
Abort
, or uncheck the
Setup Mode
checkbox and then
perform the acquisition.
Dots
This sets the maximum events to be displayed during
Setup Mode
. The default size is 1,000 events.
The
Setup Mode
rolling displays are updated up to three times per second. To change the size, edit
Содержание Navios
Страница 8: ...773232AH viii Safety Notice Safety Symbols ...
Страница 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Страница 38: ...773232AH xxxviii Introduction Graphics ...
Страница 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Страница 72: ...773232AH 3 18 Operation Principles Plot Display ...
Страница 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Страница 140: ...773232AH 5 34 System Overview Using The System Help ...
Страница 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Страница 194: ...773232AH 8 16 Running Samples Data Review ...
Страница 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Страница 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Страница 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Страница 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Страница 540: ...773232AH References 2 References ...
Страница 582: ...Index 36 Index ...
Страница 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Страница 585: ......