Portable X-ray Units
Service Manual
SM-1076R2
126
7.
Remove the Control Driver Board as described in
Section 5.3.14 (steps
2. to 10.)
without disassembling the Filaments HF Board (
A9500--01
).
8.
From the HV Transformer, disconnect the J10 Ribbon cable, the wires
connected to TR1--2, TR1--3, TR1--1 (GND) and the three GND wires
fixed to the HV Transformer screws.
GND Wire
GND Wires
J10
TR1--3
TR1--2
TR1--1 (GND)
Control Driver Board
9.
Remove the Collimator Adaptation Ring.