Portable X-ray Units
Service Manual
SM-1076R2
69
ERROR CODE :
E48
DESCRIPTION
:
Bucky / Digital Panel interface error.
The user is trying to perform an exposure with a Bucky or a Digital Panel but this
interface is not available.
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The workstation configuration is wrong.
The unit has not installed the proper software for Bucky / Digital Panel interface.
ACTIONS
Configure the workstation without bucky/digital interface or ask for the proper software
which supports the bucky/digital interface.
1.
If the error is due to a wrong configured Workstation, perform actions described
in error codes “E02” and “E03”. If a Bucky / Digital Panel is not interfaced to this
Workstation, configure it for “
Direct
” Workstation.
2.
If the error is due to a wrong software version, contact to the manufacturer.
ERROR CODE :
E50
DESCRIPTION
:
Exposure has been aborted by the operator or defective exposure device. The user has
released the exposure device/s before the end of the exposure (Prep and/or Exp).
“Recoverable” with the “
Reset
” push-button.
POSIBLE CAUSES
The user has prematurely released the exposure device.
Defective exposure device (Handswitch, Control Panel, ...).
ACTIONS
1.
Check the correct status of the exposure device (Handswitch, Control Panel, ...)
and their respective cable connections.
2.
Verify the possible defective hardware related to these inputs and replace the
defective hardware, if needed.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...