Portable X-ray Units
Service Manual
SM-1076R2
55
ERROR CODE :
E23
DESCRIPTION
:
Error while writing in the EEPROM. “Unrecoverable”, the system remains lock.
POSIBLE CAUSES
The EEPROM is defective.
ACTIONS
1.
Initialize the EEPROM (
refer to Section 5.1.1
).
2.
If the error persists, replace the Portable Control Board (
A3175--03
) and initialize
the new EEPROM (
refer to Section 5.1.1
).
ERROR CODE :
E24
DESCRIPTION
:
Bucky / Digital Panel error. A Workstation configured as Bucky / Digital Panel has
elapsed 10 seconds since exposure signal was activated while the equipment was
”
Ready
” without any acknowledge from the Bucky or Digital Panel. “Recoverable” with
the “
Reset
” push-button.
POSIBLE CAUSES
Defective interconnection cable or hardware related to the signals RX_RQ (Request) or
RX_ORDER (Acknowledge). Or the Bucky / Digital Panel is not sending the RX_ORDER
signal.
ACTIONS
If the error persists, check the cable connections and signal of the Image Receptor, or
replace the defective hardware.
1.
Check cable connections and signal from/to the Image Receptor .
2.
Check the RX signals in:
G
connector J9.3 (RX_RQ), J9.4 (RX_ACKN) on Portable Control Board
(
A3175--03
).
G
connector J1.9 (RX_RQ), J1.4 (RX_ACKN) of RS--232 connector of the
Portable Unit.
G
connector J5.2 and J1.10 (RX_RQ), and J5.3 and J1.23 (RX_ACKN) on
the Remote Console Connectors (
A3632--02
) (only for Portable Units with
Trolley).
3.
Verify the possible defective hardware related to these inputs.
4.
Replace the defective hardware.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...