Portable X-ray Units
Service Manual
SM-1076R2
91
4.
Cut the Tie Warps when needed. Cut the round connector of the TA cable
and take it out through the red (round) Current Transformer and take out
TB cable (it is not necessary to cut this one) through the grey (square)
Voltage Measurer.
Voltage Measurer
Current Transformer
5.
The Control Driver Board assembly comprises this Board, the IGBT, an
external Resistor and an aluminium Shunt Plate that is fixed to the HV
Transformer with two screws. Unscrew the two (2) allen screws located
under both sides of the Control Driver Board and take out the assembly.
Shunt Plate Screws