Portable X-ray Units
Service Manual
SM-1076R2
9
2.2.2
CONFIGURATION OF mA STATIONS FOR EACH FOCUS
1.
Enter Configuration menu C03.
2.
Enter sub-menu C3.1. The kVp Display shows (C3.1) and the mAs
Display shows (Factory set 20.00).
3.
To change the corresponding maximum value of mA station for Small
Focus, press
“mAs Increase”
or
“mAs Decrease”.
Factory default value
is 20 mA.
4.
Press
“Reset”
push-button to save value.
5.
Press
“kVp Increase”
, the kVp Display shows sub-menu (C3.2) and mAs
Display shows (Factory set 5.000).
6.
To change the corresponding minimum value of mA station for Large
Focus, press
“mAs Increase”
or
“mAs Decrease”.
Factory default value
is 5 mA.
7.
Press
“Reset”
push-button to save value.
8.
Press
“Collimator Light”
push-button to exit the sub-menus
.
The kVp
Display shows (C03).
Maximum mA station recommended for the ”Small Focus” is 20.0.
The minimum mA station for “Large Focus” should be the lowest
mA station allowed by the Unit. For 2 kW unit mA Station are
limited.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...