Portable X-ray Units
Service Manual
SM-1076R2
21
5.
Once the correct mA value is reached, save the value pressing
“Reset”
push-button (one beep will sound and the saved value is shown on the
kVp Display). It is important to write down this data for later reference
during calibration (Refer to Table 3-1).
Once a value for filament current is saved, the value remains in
memory for the next mA station. This feature helps to save time at
entering new data for the next station.
6.
Repeat above procedure to calibrate the Filament Current for 80 kVp at
the lowest mA station in Large Focal Spot. The Filament Current value
is few steps below the one calibrated for 40 kVp. (
Refer to Table 3-1)
.
The calibration data for Filament Current should not be repeated
at any station.
Table 3-1
Calibration Data for Reference
kVp
mA
Value
Example
40
Minimum Station
707
80
Minimum Station
702
Values above 999 are shown on kVp display as “= XX”, a.e. “=23“
means 1023.
7.
To exit Calibration Mode, press and hold
“Collimator Light”
push-button
and then press and release
“mAs Increase”
push-button (two beeps
sound). Then release
“Collimator Light”
push-button.
Note
.
Note
.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...