Portable X-ray Units
Service Manual
SM-1076R2
145
5.4.22.4 REMOTE CONSOLE 3P BOARD
1.
Access inside the 3P Console Box as described in
Section 5.4.22.1.
2.
Unplug the cables connected to J2, J3, P6 and P3 (Keyboard cable).
3.
Unscrew the six (6) nuts of the Remote Console 3P Board (A3634--01)
and remove it.
J2
J3
P6
Nuts (6)
P3
4.
Install the new Remote Console 3P Board.
5.
Reconnect all the cables and assemble in reverse order.
6.
Assemble the Unit as described in
Section 5.4.22.1
.
Summary of Contents for SPL-HF 2.0