5.4.23.6 TOUCH SCREEN
1.
Access inside the Touch Screen Console Box as described in
Section
5.4.23.1.
2.
Unscrew the four (4) nuts located at each corner of the PC Kit and
remove it. Take care with the washers.
Nuts (4)
PC Kit (includes PC and
Compact APR Base Board)
3.
Unplug the cables connected to J1, J9 and J10 of the Compact APR
Base Board to facilitate the access to the Board Support and Touch
Screen.
J1
J9
J10