Portable X-ray Units
Service Manual
SM-1076R2
149
5.4.23.2 POWER LINE CABLE
1.
Access inside the Touch Screen Console Box as described in
Section
5.4.23.1.
2.
Disconnect the GND from the Console Box GND stud.
3.
Remove the Plastic Protector Cover by unscrewing the two (2) nuts.
4.
Disconnect the Power Line wires (Faston connectors) in the rack.
5.
Unscrew the Plastic Nut of the Cable Gland.
6.
Pass the disconnected Power Line Cable through the Cable Gland.
GND
Nuts (2)
Power Line wires
Cable Gland
7.
Pass the new Power Line Cable through the Cable Gland and screw the
Plastic Nut.
8.
Connect the Power Line wires (Faston connectors) in the rack and the
GND to the Console Box GND stud.
9.
Assemble the Plastic Protector Cover by screwing the two (2) nuts.
10.
Assemble the Unit as described in
Section 5.4.23.1
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...