Portable X-ray Units
Service Manual
SM-1076R2
25
3.2.4
AUTOCALIBRATION
1.
Select Large Focal Spot.
2.
To enter Autocalibration,
press and hold
“Collimator Light”
push-button
and then press and release
“kVp Decrease”
push-button (two beeps
sound). The text (CAL Auto) appears on the displays while
“Collimator
Light”
is pressed. Then release
“Collimator Light”
push-button.
kVp
mAs
1
2
Press and hold Collimator Light
Press and release kVp Decrease
Release Collimator Light
3
3.
Double check that the Large Focal Spot Led is On. The Displays show
(YES no). Press
“kVp Increase”
push-button to select “YES” (to select
“
no
” and exit Autocalibration process, press
“mAs Increase”
push-button).
kVp
mAs
Double check the selected Focus
4.
To start the Autocalibration process, press “
Prep
” and then press and
hold
“Exposure”
with the Handswitch (or the corresponding push-buttons
on the Control Console).
Be prepared to write down the Autocalibration data in Table 3-3.
The Unit starts a series of exposures, a maximum of 20 in order to start the
process and a maximum of 10 in order to find each mA Station. The kVp Display
shows (n20) in a countdown sequence (a.e. n19, n18, n17...), or (n10) in a
countdown sequence (a.e. n9, n8, n7...) until the Unit finds all the mA Stations
for that Focal Spot. It may not be needed to perform all the exposures.
Once a Station is found, a long bip sounds and kVp display shows the
Calibration value for that mA Station. Then the process goes on with the
following mA Station, until all mA Stations are Autocalibrated. The Displays
show (End Auto).
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...