101
5.
Carefully remove the Lower Cover. Pass the Power Line cable through
the Connectors Panel window.
6.
If applicable remove the DAP Rails and SID Guard attached to the
Collimator (
Refer to Section 5.3.7
).
7.
Fully unscrew the four (4) Collimator Adjusting Screws and carefully put
the Collimator aside without disconnecting the cables.
8.
Put aside the Central Support by removing the four (4) fixing screws
located at both sides and the two (2) fixing screws located at the front side
of the Central Support.
Fixing Screw
Fixing Screw
Central Support
Fixing Screw