Portable X-ray Units
Service Manual
SM-1076R2
8
2.2.1
CONFIGURATION OF WORKSTATIONS (1) -- (2)
To select Workstation 1 press the “
Direct
” Workstation push-button, to select
Workstation 2 press the “
Receptor
” (Bucky or Digital Detector) Workstation
push-button.
In case the Receptor Workstation for the Digital Detector or Bucky
is selected and no Digital Detector or Bucky is connected to the
Unit or the connection is malfunctioning, error code “E24” appears
on the Display. Press “Reset” push-button and check connections
and signals of the Digital Detector or Bucky, or select the
Workstation for Direct if no Receptor is needed.
Factory Setting
Workstation 1
Direct
Workstation 2
Bucky or Detector
Workstation 1 or 2
if not Bucky or Detector
C 1.1
1
C 2.1
1
C 1.2
0
C 2.2
1
Set all values 0
C 1.3
0
C 2.3
1
Set all values 0
C 1.4
0
C 2.4
0
In case a Workstation configuration needs to be modified:
1.
Enter Configuration Menu C01 (for Workstation-1) or C02 (for
Workstation-2).
2.
Enter sub-menu C1.X or C2.X, the kVp Display shows (C1.X) or (C2.X)
and the mAs Display shows the selectable values.
As example, enter sub-menu C1.1 or C2.1. The kVp Display shows
(C1.1) or (C2.1) and the mAs Display shows (0) or (1). “X-ray Tube
present” value is “
0
” for Workstation not used or “
1
” for X-ray Tube--1
present (Workstation used).
3.
Press
“mAs Increase”
to change the corresponding value from a.e. “
0
”
(Workstation not used) to “
1
” (Tube present -- Workstation used) or
viceversa.
4.
Press
“Reset”
push-button to save value.
Note
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...