Portable X-ray Units
Service Manual
SM-1076R2
128
17.
Re-install the Collimator.
Collimator replacement will affect centering of Collimator
light. Screwing may modify position of Collimator and it
must be manually centered while tightening the Safety
Screws. It is recommended 5 turns to each screw, check
parallelism between the Collimator and Unit Chassis, correct
position if necessary and finally tighten each Safety Screw
with the same number of turns (one by one).
Once the Collimator is replaced, the Unit fully assembled and
ready for operation, perform the Adjustments described in
Section 4.
18.
If applicable, assemble the DAP Rails and SID Guard attached to the
Collimator.
19.
Re-install the X-ray Power Unit in the “U” shaped support of the Arm (do
not scratch the covers). The washers must be installed in the same order
that were removed and the two (2) special screws must be tighten using
a torque of 30 Nm.
Special Screws (2)
Power / Communication Harness
Special Screws (2)
“U” Shaped Support
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...