Portable X-ray Units
Service Manual
SM-1076R2
22
3.2.2
CALIBRATION OF kVp GAIN
1.
To enter Calibration Menu
, press and hold
“Collimator Light”
push-button and then press and release
“mAs Decrease”
push-button
(two beeps sound). (CAL PArA) appears on the displays while
“Collimator Light”
is pressed
.
Then release
“Collimator Light”
push-button. The kVp Display shows P01 and one of the Focal Spot Leds
blinks.
1
2
Press and hold Collimator Light
Press and release mAs Decrease
Release Collimator Light
3
mAs
kVp
2.
Press
“Collimator Light”
to enter calibration of kVp Gain (P01).
3.
Place a Non-Invasive kVp meter aligned to the X-ray beam at SID
100 cm. Open the light field to center it with the kVp meter.
4.
Make an exposure (parameters are automatically selected: 80 kVp, 5mA
(or minimum mA) and 100 ms), read kVp and take note of the value. Enter
kVp read value in the mAs Display pressing
“mAs Increase”
or
“mAs
Decrease”
push-buttons.
5.
Save the value by pressing
“Reset”
push-button. The mAs Display shows
kVp Gain value during two seconds (this value represents the Gain
correction automatically applied, 200 is the reference value).
6.
Press
“Collimator Light”
to exit menu P01.
7.
To exit Calibration Mode, press and hold “
Collimator Light”
push-button
and then press and release
“mAs Decrease”
(two beeps sound). Then
release
“Collimator Light”
(kVp and mAs parameters are shown).
kVp
kVp
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...