7.
kVp Test:
G
Place a Non-Invasive kVp Meter aligned to the X-ray beam at SID
100 cm to measure kVp. Open the light field to center it with the
kVp Meter.
G
Select 80 kVp and the lowest mA station in Large Focus. Make an
exposure and note the kVp read on the Non-invasive kVp Meter
at the end of the exposure.
G
Check that the kVp value read must be the selected on the Control
Panel: 80
±
(3% +1 kVp). If the kVp value does not comply with
it, perform the respective Calibration procedures (
refer to
Section 3
).
8.
mA Test:
G
Select 80 kVp, Small Focus and the highest mA station in Small
Focus (read the mA station on the Control Panel by pressing any
“
Focal Spot
” and “
mAs Decrease
” push-buttons at the same time).
Make an exposure and measure the mA signal value with the
Oscilloscope. Verify that the signal is flat and noiseless.
G
Select 80 kVp, Large Focus and the lowest mA station in Large
Focus (read the mA station on the Control Panel by pressing any
“
Focal Spot
” and “
mAs Decrease
” push-buttons at the same time).
Make an exposure and measure the mA signal value with the
Oscilloscope. Verify that the signal is flat and noiseless.
G
Check that the mA values read must be the selected on the Control
Panel
±(
4% + 1 mA). If any mA value does not comply with it,
perform the respective Calibration procedures (
refer to Section 3
).
9.
Turn the Unit OFF. Disconnect the Oscilloscope and re-install the Upper
Cover of the Unit.
6.7
ALIGNMENT OF THE X-RAY BEAM AND FIELD SIZE INDICATOR TEST
Check the “
Alignment of the X-ray Beam
” and the “
Field Size Indicator Test
” as
described in
Section 4.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...