137
5.4.18
WHEEL GREENTYRE FUSION
1.
Place the Arm in Parking Position. To place the Arm in the Parking
Position:
a.
Stand up in front of the Unit.
b.
Position the Power Module in parallel to the Arm.
c.
Unlock the Gas Spring with the Arm Lock Lever.
d.
Push the Power Module downwards with both hands until the
parking detent is locked.
2.
Lay the Trolley down on its side with the leg fully extended.
3.
Unscrew the nut at the upper wheel and replace it.
Nut
4.
Lay the Trolley down on the other side and replace the other wheel.