Portable X-ray Units
Service Manual
SM-1076R2
104
5.3.16.2 CIRCUIT BREAKER
1.
Access inside the Central Support as described in
Section 5.3.16.1
.
2.
Remove the fixing screws located at the back of the Central Support.
3.
Cut the Tie Wrap which secures the two wires connected to the Power
Supply. Disconnect the four wires attached keeping in mind the
correspondence of each cable.
4.
Loosen without removing the two nuts which fixes the two metallic plates
in contact with the Circuit Breaker and the EMC Input Filter Board.
Disconnect the two metallic plates from the Circuit Breaker, do not
remove them from the EMC Input Filter Board.
5.
Pull the two Latches and remove the Circuit Breaker from its DIN Rail.
Power Line Wires
Wires (2) to Power Supply
Latch
DIN Rail
Tie Wrap
Nuts (2) of Metallic Plates
on EMC Input Filter Board
6.
Replace the Circuit Breaker, reconnect in reverse order and assemble
the Unit as described in
Section 5.3.16.1
.
Summary of Contents for SPL-HF 2.0
Page 6: ...Portable X ray Units Service Manual...
Page 42: ...Portable X ray Units Service Manual SM 1076R2 32 This page intentionally left blank...
Page 48: ...Portable X ray Units Service Manual SM 1076R2 38 This page intentionally left blank...
Page 166: ...Portable X ray Units Service Manual SM 1076R2 156 This page intentionally left blank...
Page 176: ...Portable X ray Units Service Manual SM 1076R2 166 A3 A3 A4 A6 A7 A8 A9 A10 A11 A12 A13...
Page 177: ...Portable X ray Units Service Manual SM 1076R2 167 A22 A21 A20 A23 A15 A14 A24...
Page 180: ...Portable X ray Units Service Manual SM 1076R2 170 B10 B1 B2 B4 B5 B9 B11 B12 B13 B15...
Page 184: ...Portable X ray Units Service Manual SM 1076R2 174 C1 C2 C3 C4 C5 C6 C7 C8 C9 C10 C11 C13 C14...
Page 185: ...Portable X ray Units Service Manual SM 1076R2 175 C15 C16 C17 C18 C19...
Page 190: ...Portable X ray Units Service Manual SM 1076R2 180 This page intentionally left blank...